Automated Fluorescence Lifetime Spectrometer "Easy Tau" |
FluoTime 300 |
Specifications
| System |
| Optical configuration |
L-Geometry |
| Mode of operation |
Time-Correlated Single Photon Counting (TCSPC), Multichannel Scaling (MCS) |
 |
| Monochromator |
| Type |
Czerny-Turner design |
Czerny-Turner design |
| Focal length |
150 mm |
300 mm |
| Aperture |
F/4.2 |
F/3.9 |
| Stray light rejection |
10-5 |
10-5 |
Standard grating (others on request) |
1200 g/mm, blazed at 500 nm |
1200 g/mm, blazed at 500 nm |
| Resolution |
0.4 nm (at 438.5 nm, slit width 10 µm, 1200 g/mm grating) |
0.1 nm (at 438.5 nm, slit width 10 µm, 1200 g/mm grating) |
| Step size (min) |
0.01 nm |
0.01 nm |
| Slit width adjustable between |
10 µm - 4 mm (continuously adjustable, completely motorized) |
10 µm - 4 mm (continuously adjustable, completely motorized) |
| Wavelength accuracy |
0.25 nm (1200 g/mm grating) |
0.2 nm (1200 g/mm grating) |
| Wavelength repeatability |
0.1 nm |
0.1 nm |
 |
| Excitation Sources |
| Light source |
Pulsed LEDs (PLS) |
|
Laser Diode Heads (LDH) |
|
pulsed Xenon lamp |
|
cw Xenon lamp |
| Wavelengths |
245 - 600 nm |
375 - 510, 530, 595 635 - 1550 nm |
200 - 900 nm |
200 - 900 nm |
| Pulse width |
400 ps - 1 ns |
60 - 500 ps |
<1 µs |
--- |
| Repetition rate |
up to 40 MHz |
up to 40 MHz |
up to 300 Hz |
--- |
|
(80 MHz optional) |
 |
| Detectors |
| Type |
PMT (PMA-C) |
|
MCP-PMT |
|
NIR-PMT |
| Spectral range in nm |
185 - 820 |
300 - 900 |
160 - 650 |
160 - 850 |
950 - 1400 |
| Dark counts (at 20°C, typ. value) |
< 150 cps |
< 900 cps |
< 50 cps |
< 500 cps |
< 10.000 cps |
|
other detectors types available on request |
 |
| Data Acquisition |
| Type |
PicoHarp 300 |
|
NanoHarp 250 |
| Number of time channels/curve |
up to 65536 |
up to 262144 |
| Count depth |
16 bit |
18 bit |
| Time resolution (bin width) |
4 ps |
4 ns or 32 ns |
| Differential non-linearity |
< 0.1 % rms, < 5 % pp. |
< 0.1 % rms, < 1 % pp. |
| Dead Time |
< 95 ns |
8 ns |
| Discriminator level, zero cross adjust |
software adjustable |
| Collection time |
1 ms - 10 h |
 |
| Data Analysis Software |
| Type |
FluoFit (Professional) |
| Analysis possibilities |
exponential decay, lifetime distribution, anisotropy, global analysis |
 |
| Operation Environment |
| Computer system |
Windows XP/Vista/7 |
| Power requirements |
110 to 230 V 50/60 Hz |
| Dimensions (base unit) |
| without steady-state option |
900 mm × 550 mm × 400 mm (w × d × h) |
| with steady-state option |
900 mm × 1100 mm × 400 mm (w × d × h) |
All Information given here is reliable to our best knowledge. However, no responsibility is assumed for possible inaccuracies or omissions. Specifications and external appearances are subject to change without notice. © PicoQuant GmbH, 2012
|