Single Photon Counting Module τ-SPAD

Specifications at 25°C

Spectral range 400 nm - 1100 nm
Active area diameter 150 µm
Photon timing resolution* 350 ps - 800 ps (FWHM, depending on module, wavelength and signal rate)
Afterpulsing probability (0 to 500 ns)* < 1 % (typical)
Dead time < 60 ns (typical)
Fiber connector type (optional) FC/PC
Dark count rate
τ-SPAD-20 < 20 cps (available upon request)
τ-SPAD-50 < 50 cps
τ-SPAD-100 < 100 cps
Photon detection efficiency* (typical values, without fiber connector)
@ 405 nm 20 %
@ 470 nm 45 %
@ 670 nm 75 %
@ 890 nm 35 %
Losses due to fiber connector: approx. 10%
Input / Output
NIM output
Pulse width 15 ns - 25 ns
Pulse amplitude -0.8 V to -1 V (into 50 Ohms)
Connector type SMA
TTL output
Pulse width 15 ns - 25 ns
Pulse amplitude > 2.4 V (into 50 Ohms)
Connector type Lemo, type EPS.00.250
Gating input
Input voltage TTL control, TTL high (>2.4 V) enables counting
Response time disable: < 40 ns (typ. 20 ns); enable: < 100 ns (typ. 85 ns)
Connector type SMA
Operating conditions
Supply voltage 12 V
Supply current 1.5 A (at turn on), 0.3 A (at 1 Mcps)
Operating temperature 10 °C - 40 °C
* measured by illuminating < 30 µm in the center of the active area

Typical detection efficiency vs. wavelength (without fiber connector)

Typical detection efficiency of the Tau-SPAD

All Information given here is reliable to our best knowledge. However, no responsibility is assumed for possible inaccuracies or omissions. Specifications and external appearances are subject to change without notice.
© PicoQuant GmbH, 1998 - 2012


Home PicoQuant GmbH
Rudower Chaussee 29 (IGZ)
12489 Berlin
Germany
Phone:  ++4930 6392 6929
Fax: ++4930 6392 6561
Email: infopicoquant.com